APPENDIX A-5: DAS Triggers and Parameters Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
A-32
Table A-6: T300/T300M and M300E/EM DAS Parameters, Revision L.8
Name Description Units
DETMES Detector measure reading mV
DETREF Detector reference reading mV
RATIO M/R ratio. none
SLOPE1 Slope for range #1 none
SLOPE2 Slope for range #2 none
OFSET1 Offset for range #1 none
OFSET2 Offset for range #2 none
CO2SLP
5
CO
2
slope none
CO2OFS
5
CO
2
offset %
O2SLPE
8
O
2
slope none
O2OFST
8
O
2
offset %
AZERO
1,2
Auto-zero reading M/R
ZSCNC1
Concentration for range #1 during zero/span calibration, just before
computing new slope and offset
PPM
ZSCNC2
Concentration for range #2 during zero/span calibration, just before
computing new slope and offset
PPM
CO2ZSC
5
CO
2
concentration during zero/span calibration, just before
computing new slope and offset
%
O2ZSCN
8
O
2
concentration during zero/span calibration, just before computing
new slope and offset
%
CONC1 Concentration for range #1 PPM
CONC2 Concentration for range #2 PPM
CO2CNC
5
CO
2
concentration %
O2CONC
8
O
2
concentration %
STABIL Concentration stability PPM
BNTEMP Bench temperature
C
BNCDTY Bench temperature control duty cycle Fraction
(0.0 = off,
1.0 = on full)
WTEMP Wheel temperature
C
WHLDTY Wheel temperature control duty cycle Fraction
(0.0 = off,
1.0 = on full)
O2TEMP
8
O
2
sensor cell temperature
C
SMPTMP Sample temperature
C
SMPFLW
6
Sample flow cc/m
SMPPRS Sample pressure "Hg
VACUUM
1, 3, 6
Vacuum pressure "Hg
PRGPRS
4
Purge pressure PSIG
BOXTMP Internal box temperature
C
BX2TMP
2
,
OVNTMP
9
Internal box temperature #2/oven
C
04188D DCN5752