Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840) APPENDIX A-2: Setup Variables For Serial I/O
A-19
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
3000 (not
used)
REF_SOURCE_LIMIT mV
Warnings:
1100–4800,
25–4800
3, 4, 15
1–5000
Reference source warning limits.
Set point is not used.
FACTORY_OPT BitFlag 512,
768
5
0–65535
Factory option flags. Add values
to combine flags.
1 = enable dilution factor
2 = zero/span valves installed
4 = enable conc. alarms
8 = enable linearity adjustment
factor
16 = display units in
concentration field
32 = enable software-controlled
maintenance mode
64
3, 5
= span valve installed
128 = enable switch-controlled
maintenance mode
256 = compute only offset during
zero calibration
512 = 220 V A/C power
1024 = non-zero offset
calibration (linearity adjustment
must also be enabled)
2048 = enable Internet option
7
4096 = use “old” style numeric
data entry menus when editing
conc. table
8192 = locate high range and
zero cal. status outputs on relays
04188D DCN5752