Troubleshooting & Repair Teledyne API – Technical Manual - Model 300E Family CO Analyzers
296
13.5.7.4. Status Outputs
The procedure below can be used to test the Status outputs:
1. Connect a jumper between the “D“ pin and the “” pin on the status output connector.
2. Connect a 1000 ohm resistor between the “+” pin and the pin for the status output that is being tested.
3. Connect a voltmeter between the “” pin and the pin of the output being tested (see table below).
Under the DIAG SIGNAL I/O menu (see Section 13.1.3), scroll through the inputs and outputs until you get to
the output in question. Alternately turn on and off the
output noting the voltage on the voltmeter, it should vary
between 0 volts for ON and 5 volts for OFF.
Table 13-12: Status Outputs Check
PIN (LEFT TO RIGHT) STATUS
1 SYSTEM OK
2 CONC VALID
3 HIGH RANGE
4 ZERO CAL
5 SPAN CAL
6 DIAG MODE
7 SPARE
8 SPARE