Teledyne M300EM Carbon Monoxide Alarm User Manual


 
Advanced Features Teledyne API – Technical Manual - Model 300E Family CO Analyzers
112
7.1.5. SETUP DAS EDIT: ACCESSING THE IDAS EDIT MODE
iDAS configuration is most conveniently done through the APICOM remote control program. The following list of
key strokes shows how to edit the iDAS using the front panel.
When editing the data channels, the top line of the display indicates some of the configuration parameters.
For example, the display line:
0) CONC: ATIMER, 1, 800
Translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: One parameter is included in this channel
EVENT: This channel is set up to store 800 records.
To edit the name of a data channel, follow the above key sequence and refer to Section 7.1.5.1:
04288D DCN5752