An example of resetting the span factor is outlined below:
Certified check cell of 2880 ppm is inserted into the check cell holder
with the analyzer in Mode 6 and a reading taken. This reading is 3240.
Due to the value of the check cell exceeding 999 ppm a suitable
pathlength needs to be used to reduce the effective value of the check
cell. In this case a pathlength of 4 m is chosen, thus the effective value
of the check cell in 720 ppm (2880/4).
However, in Mode 6 it reads higher (3240); this value is effectively 810
ppm at a 4 m pathlength.
810 is keyed in as the first value in the span factor option.
The span factor is then adjusted until the second value reads 720.
Now press Enter to set the span factor.