Emerson IM-106-5500 Carbon Monoxide Alarm User Manual


 
Instruction Manual
IM-106-5500, Original Issue
August 2005
3-19
CCO 5500
After the calibration has been conducted, a CAL FACTOR will be
calculated; this determines the calibration of the instrument. The cal
factor will be displayed for a few seconds after a calibration has been
conducted, and it can also be interrogated from Mode 4 (diagnostic
mode). Enter the cal factor into the table below, as a record of
instrument operation.
Table 3-2. Calibration Data
NOTE
Factory settings were obtained under the following conditions:
6.5 ft (2 m) path length
Clean conditions
240V, 50 Hz supply
CURRENT OUTPUT
CALIBRATION
NOTE
This is set at the factory and should not be altered without due consideration.
The current output should now be set up using a calibrated milliammeter set
to DC current, 20 mA max. Conduct this procedure as follows :
1. Connect the milliammeter to the output terminals within the signal
processor - terminals +mA and -mA.
2. Enter the SET UP MODE - CONFIGURE OUTPUT - SET ZERO
OPTION, and adjust the level using the ARROW keys until 0mA is
recorded. Record the value in brackets on the display in Table 2 -
Calibration Data.
3. Enter the SETSPAN option and adjust the level using the ARROW keys
until 20 mA is recorded. Record the value in brackets on the display in
Table 2 - Calibration Data.
Calibration Data
Detector Outputs Factory Config
D2
D1
E1
E2
Calibration Data
Cal factor
Span factor
Output Calibration
Set zero
Set span