Normalization All of the normalization inputs and parameters
are set up from this mode. Press ENTER to
access the mode and the ARROW keys will
select which of the normalizing inputs are to be
changed, they are:
1. Temperature
2. Oxygen
3. Pressure
4. Water Vapor
After selecting the normalizing parameter, the user may set the standard
levels to which the measurement is to be normalized, and how the instrument
reads the value, i.e. fixed keypad input, 4-20 mA input, or on the serial data
line. Figure 4-5 illustrates the program tree for entering the normalization
parameters.
Figure 4-5. Normalization
Options
5 NORMALIZATION
Press Enter
on Exit option
Temperature Set std. levels
Set values
Exit
Value to which parameter is to be normalized
Analog I/P - use a 4-20mA input
Keypad I/P - set a fixed value from the keypad
Serial I/P - use the serial data line for this input
Exit
Oxygen Set std. levels
Set values
Exit
Value to which parameter is to be normalized
Analog I/P - WET or DRY gas, use a 4-20mA input
Keypad I/P - set a fixed value from the keypad
Serial I/P - use the serial data line for this input
Exit
Pressure Set std. levels
Set values
Exit
Value to which parameter is to be normalized
Analog I/P - use a 4-20mA input
Keypad I/P - set a fixed value from the keypad
Serial I/P - use the serial data line for this input
Exit
Water Vapor Set std. levels
Set values
Exit
Value to which parameter is to be normalized
Keypad I/P - set a fixed value from the keypad
Serial I/P - use the serial data line for this input
Exit
Press Enter
on Exit option