Teledyne API – Technical Manual - Model 300E Family CO Analyzers Optional Hardware and Software
65
5.6.2.1. Internal Pneumatics (OPT 50A)
Figure 5-3: Internal Pneumatic Flow OPT 50A – Zero/Span Valves
Table 5-1: Zero/Span Valve Operating States for Option 52
MODE VALVE CONDITION
Sample/Cal Open to SAMPLE inlet
SAMPLE
(Normal
State)
Zero/Span Open to IZS inlet
Sample/Cal Open to ZERO/SPAN valve
ZERO CAL
Zero/Span Open to IZS inlet
Sample/Cal Open to ZERO/SPAN valve
SPAN CAL
Zero/Span Open to PRESSURE SPAN inlet
04288D DCN5752