Teledyne M300E Carbon Monoxide Alarm User Manual


 
Remote Operation Teledyne API – Technical Manual - Model 300E Family CO Analyzers
152
8.1.4. COMM PORT COMMUNICATION MODES
Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in Table
8-1. As modes are selected, the analyzer sums the mode ID numbers and displays this combined number on
the front panel display. For example, if quiet mode (01), computer mode (02) and Multi-Drop-Enabled mode (32)
are selected, the analyzer would display a combined MODE ID of 35.
Table 8-1: COMM Port Communication Modes
MODE
1
ID DESCRIPTION
QUIET
1
Quiet mode suppresses any feedback from the analyzer (such as warning messages) to
the remote device and is typically used when the port is communicating with a computer
program where such intermittent messages might cause communication problems.
Such feedback is still available but a command must be issued to receive them.
COMPUTER
2
Computer mode inhibits echoing of typed characters and is used when the port is
communicating with a computer operated control program.
HESSEN
PROTOCOL
16
The Hessen communications protocol is used in some European countries. T-API P/N
02252 contains more information on this protocol.
E, 8, 1
8192
When turned on this mode switches the COMM port settings from
NO PARITY; 8 data bits; 1 stop bit to EVEN PARITY; 8 data bits; 1 stop bit.
E, 7, 1
2048
When turned on this mode switches the COM port settings from
NO PARITY; 8 DATA BITS; 1 stop bit to EVEN PARITY; 7 DATA BITS; 1 stop bit.
RS-485
1024
Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence
over multidrop mode if both are enabled.
SECURITY
4
When enabled, the serial port requires a password before it will respond (see Section
8.1.7.5). If not logged on, the only activ
e command is the '?' request for the help
screen.
MULTIDROP
PROTOCOL
32
Multidrop protocol allows a multi-instrument configuration on a single communications
channel. Multidrop requires the use of instrument IDs.
ENABLE
MODEM
64
Enables to send a modem initialization string at power-up. Asserts certain lines in the
RS-232 port to enable the modem to communicate.
ERROR
CHECKING
2
128
Fixes certain types of parity errors at certain Hessen protocol installations.
XON/XOFF
HANDSHAKE
2
256
Disables XON/XOFF data flow control also known as software handshaking.
HARDWARE
HANDSHAKE
8
Enables CTS/RTS style hardwired transmission handshaking. This style of data
transmission handshaking is commonly used with modems or terminal emulation
protocols as well as by Teledyne Instrument’s APICOM software.
HARDWARE
FIFO
2
512
Disables the HARDWARE FIFO (First In – First Out). When FIFO is enabled it improves
data transfer rate for that COM port.
COMMAND
PROMPT
4096
Enables a command prompt when in terminal mode.
1
Modes are listed in the order in which they appear in the
SETUP MORE COMM COM[1 OR 2] MODE menu
2
The default setting for this feature is ON. Do not disable unless instructed to by Teledyne API’s Customer Service
personnel.
Note
Communication Modes for each COMM port must be configured independently.