Teledyne M300E Carbon Monoxide Alarm User Manual


 
Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840) APPENDIX A-2: Setup Variables For Serial I/O
A-13
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
RANGE_MODE — SNGL
0
SNGL,
DUAL,
AUTO
Range control mode.
CONC_RANGE1 Conc. 50,
200
6
,
500
3, 8
0.1–50000 D/A concentration range 1.
CONC_RANGE2
1
Conc. 50,
200
6
,
500
3, 8
0.1–50000 D/A concentration range 2.
CO2_RANGE
10
% 15 0.1–500,
0.1–2000
16
CO
2
concentration range.
O2_RANGE
14
% 100 0.1–500 O
2
concentration range.
RS232_MODE BitFlag 0 0–65535
RS-232 COM1 mode flags. Add
values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable hardware
handshaking
16 = enable Hessen protocol
11
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF
support
512 = disable hardware FIFOs
1024 = enable RS-485 mode
2048 = even parity, 7 data bits, 1
stop bit
4096 = enable command prompt
8192 = even parity, 8 data bits, 1
stop bit
16384 = enable dedicated
MODBUS ASCII protocol
32678 = enable dedicated
MODBUS RTU or TCP protocol
16384 = enable TAI protocol
17
BAUD_RATE — 115200
0
300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM1 baud rate.
04188D DCN5752