Teledyne M300E Carbon Monoxide Alarm User Manual


 
APPENDIX A-7: MODBUS Register Map Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
A-38
MODBUS
Register Address
(dec., 0-based)
Description Units
MODBUS Coil Registers
(single-bit; read/write)
0 Maps to relay output signal 36 (MB_RELAY_36 in signal I/O list)
1 Maps to relay output signal 37 (MB_RELAY_37 in signal I/O list)
2 Maps to relay output signal 38 (MB_RELAY_38 in signal I/O list)
3 Maps to relay output signal 39 (MB_RELAY_39 in signal I/O list)
20
4
Triggers zero calibration of range #1 (on enters cal.; off exits cal.)
21
4
Triggers span calibration of range #1 (on enters cal.; off exits cal.)
22
4
Triggers zero calibration of range #2 (on enters cal.; off exits cal.)
23
4
Triggers span calibration of range #2 (on enters cal.; off exits cal.)
1
Optional
2
O
2
option
3
CO
2
option
4
Set DYN_ZERO or DYN_SPAN variables to ON to enable calculating new slope or offset. Otherwise a calibration check
is performed.
04188D DCN5752