Teledyne M300E Carbon Monoxide Alarm User Manual


 
Basic Operation Teledyne API – Technical Manual - Model 300E Family CO Analyzers
100
6.6.3.1. RNGE MODE SNGL: Configuring the M300E/EM Analyzer for SINGLE Range Mode
NOTE
This is the default reporting range mode for the analyzer.
When the single range mode is selected (SNGL), all analog CO concentration outputs (A1 and A2) are slaved
together and set to the same reporting range limits (e.g. 500.0 ppb). The span limit of this reporting range can be
set to any value within the physical range of the analyzer.
Although both outputs share the same concentration reporting range, the electronic signal ranges of the analog
outputs may still be configured for different values (e.g. 0-5 VDC, 0-10 VDC, etc; see Section 7.4.2)
To select
SNGL range mode and to set the upper limit of the range, press:
04288D DCN5752