APPENDIX A-2: Setup Variables For Serial I/O Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
A-12
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
CO_TARG_ZERO1 Conc. 0
-100.00–
999.99
Target CO concentration during
zero offset calibration of range 1.
CO_TARG_MID1_1 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #1 calibration of range
1.
CO_TARG_MID2_1 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #2 calibration of range
1.
CO_SPAN1 Conc. 40,
400
3, 8
0.01–9999.99
Target CO concentration during
internal span calibration of range
1.
CO_SLOPE1 — 1 0.001–999.999 CO slope for range 1.
CO_OFFSET1 — 0 -10–10 CO offset for range 1.
CAL_BOX_TEMP1 ºC 30 0–100
Calibrated box temperature for
range 1.
CO_TARG_ZERO2 Conc. 0
-100.00–
999.99
Target CO concentration during
zero offset calibration of range 2.
CO_TARG_MID1_2 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #1 calibration of range
2.
CO_TARG_MID2_2 Conc. 50
5
,
300
0.01–9999.99
Target CO concentration during
mid-point #2 calibration of range
2.
CO_SPAN2 Conc. 40,
400
3, 8
0.01–9999.99
Target CO concentration during
internal span calibration of range
2.
CO_SLOPE2 — 1 0.001–999.999 CO slope for range 2.
CO_OFFSET2 — 0 -10–10 CO offset for range 2.
CAL_BOX_TEMP2 ºC 30 0–100
Calibrated box temperature for
range 2.
CO2_TARG_MID1_CONC
10
% 6,
800
16
0.1–1000,
0.1–2000
16
Target CO
2
concentration during
mid-point #1 calibration.
CO2_TARG_MID2_CONC
10
% 6,
800
16
0.1–1000,
0.1–2000
16
Target CO
2
concentration during
mid-point #2 calibration.
CO2_TARG_SPAN_CON
C
10
% 12 0.1–1000,
0.1–2000
16
Target CO
2
concentration during
span calibration.
CO2_SLOPE
10
1 0.5–5 CO
2
slope.
CO2_OFFSET
10
% 0 -10–10,
-100–100
16
CO
2
offset.
O2_TARG_SPAN_CONC
14
% 20.95 0.1–100
Target O
2
concentration during
span calibration.
O2_SLOPE
14
1 0.5–2 O
2
slope.
O2_OFFSET
14
% 0 -10–10 O
2
offset.
04188D DCN5752