Teledyne M300E Carbon Monoxide Alarm User Manual


 
Maintenance Schedule & Procedures Teledyne API – Technical Manual - Model 300E Family CO Analyzers
266
12.3.4. PERFORMING A SAMPLE FLOW CHECK
CAUTION
GENERAL SAFETY HAZARD
Always use a separate calibrated flow meter capable of measuring flows in the 0 – 1000
cm
3
/min range to measure the gas flow rate though the analyzer.
DO NOT use the built in flow measurement viewable from the Front Panel of the
instrument. This measurement is only for detecting major flow interruptions such as
clogged or plugged gas lines.
See Figure 3-2 for sample port location.
1. Attach the Flow Meter to the sample inlet port on the rear panel. Ensure that the inlet to the Flow Meter
is at atmospheric pressure.
2. Sample flow should be 800 cm
3
/min 10%.
3. Once an accurate measurement has been recorded by the method described above, adjust the
analyzer’s internal flow sensors (See Section 9.6.3).
Low flow
s indicate blockage somewhere in the pneumatic pathway, typically a plugged sintered filter or critical
flow orifice in one of the analyzer’s flow control assemblies. High flows indicate leaks downstream of the Flow
Control Assembly.
12.3.5. CLEANING THE OPTICAL BENCH
The M300E/EM sensor assembly and optical bench are complex and delicate. Disassembly and cleaning is not
recommended. Please check with the factory before disassembling the optical bench.
12.3.6. CLEANING EXTERIOR SURFACES OF THE M300E/EM
If necessary, the exterior surfaces of the M300E/EM can be cleaned with a clean damp cloth. Do not submerge
any part of the instrument in water or cleaning solution.
04288D DCN5752