6.7.2.5. Editing DAS Parameters
Data channels can be edited individually from the front panel without affecting other data channels.
However, when editing a data channel, such as during adding, deleting or editing parameters, all data for
that particular channel will be lost, because the DAS can store only data of one format (number of
parameter columns etc.) for any given channel. In addition, an DAS configuration can only be uploaded
remotely as an entire set of channels. Hence, remote update of the DAS will always delete all current
channels and stored data.
To modify, add or delete a parameter, follow the instruction shown in Section 6.7.2.2 then press:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X PARAMETERS:1
<SET SET> EDIT PRINT EXIT
SETUP X.X 0) PARAM=CONC1, MODE=AVG
PREV NEXT INS DEL EDIT EXIT
Press SET> button until…
SETUP X.X EDIT PAR
A
MS (DELETE DATA)
YES NO
NO returns to
the previous
menu and
retains all data.
Moves the
display between
existing
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
Deletes the Parameter
currently displayed.
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
YES will delete
all data in that
entire channel.
From the DATA ACQUISITION menu
(see Section 6.7.2.2)