Operating Instructions Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 78
6.7.2.3. Editing DAS Data Channels
DAS configuration is most conveniently done through the APICOM remote control program. The
following list of key strokes shows how to edit using the front panel.
Main Data Acquisition Menu
Edit Data Channel Menu
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X 0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
EXIT will return to the
previous SAMPLE
display.
SETUP X.X DATA ACQUISITION
VIEW
EDIT EXIT
Moves the
display up &
down the list of
Data Channels
Inserts a new Data
Channel into the list
BEFORE the Channel
currently being displayed
Deletes The Data
Channel currently
being displayed
Exports the
configuration of all
data channels to
RS-232 interface.
Exits to the Main
Data Acquisition
Menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRNT EXIT
Moves the display
between the
PROPERTIES for this
data channel.
Reports the configuration of current
data channels to the RS-232 ports.
Exits returns to the
previous Menu
Allows to edit the channel name, see next key sequence.
When editing the data channels, the top line of the display indicates some of the configuration
parameters. For example, the display line:
0) CONC: ATIMER, 4, 800
translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.