Teledyne gfc 7000ta Carbon Monoxide Alarm User Manual


 
Operating Instructions Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 98
Example: If all passwords are enabled, the following keypad sequence would be required to enter the
SETUP menu:
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SAMPLE
ENTER SETUP PASS: 0
0 0 0 ENTR EXIT
prompts for password
number
See Table 6-8 for
Passwords and Levels
Example: this
password enables the
SETUP mode
SAMPLE ENTER SETUP PASS: 0
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
Press individual
buttons to set
numbers
Note that the instrument still prompts for a password when entering the VARS and DIAG menus, even if
passwords are disabled, but it displays the default password (818) upon entering these menus. The user
only has to press ENTR to access the password-protected menus but does not have to enter the required
number code.