Teledyne gfc 7000ta Carbon Monoxide Alarm User Manual


 
Operating Instructions Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 85
6.7.2.8. RS-232 Report Function
The GFC 7000TA DAS can automatically report data to the communications ports, where they can be
captured with a terminal emulation program or simply viewed by the user.
To enable automatic COM port reporting, follow the instruction shown in Section 6.7.2.2 then press:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X RS-232 REPORT: OFF
<SET SET> EDIT PRINT EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X RS-232 REPORT: OFF
OFF ENTR EXIT
Toggle to turn
reporting ON or OFF
Press SET> key until…
From the DATA ACQUISITION menu
(see Section 6.7.2.2)
6.7.2.9. Compact Report
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting
each parameter in one channel on a separate line, up to five parameters are reported in one line.
6.7.2.10. Starting Date
This option allows to specify a starting date for any given channel in case the user wants to start data
acquisition only after a certain time and date. If the Starting Date is in the past, the DAS ignores this
setting.