Teledyne gfc 7000ta Carbon Monoxide Alarm User Manual


 
Operating Instructions Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 122
To enter the signal I/O test mode, press:
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
EXAMPLE
SAMPLE ENTER SETUP PASS: 81
8
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG I / O Test Signals Displayed Here
PREV NEXT JUMP PRNT EXIT
Use JUMPto go
directly to a specific
signal location
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O JUMP TO: 12
1 2 ENTR EXIT
EXAMPLE:
Enter 12 to Jump to
12) ST_CONC_VALID
DIAG I / O ST_CONC_VALID = ON
PREV NEXT JUMP ON PRNT EXIT
Exit to return
to the
DIAG menu
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
Use NEXT &PREV to
move between signal
types.
Pressing PRNT will send a formatted printout to the serial port and can be captured
with a computer or other output device.
EXIT returns
to the main
SAMPLE display
6.13.3. Analog Output Step Test
This test can be used to check the accuracy and proper operation of the analog outputs. The test forces
all four analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20%
increments. This test is useful to verify the operation of the data logging/recording devices attached to
the analyzer.