Teledyne gfc 7000ta Carbon Monoxide Alarm User Manual


 
Operating Instructions Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 92
6.8.3. Single Range Mode (SNGL)
This is the default reporting range mode for the analyzer. In single range mode both A1 and A2 are set to
the same reporting range. This reporting range can be any value between 50 ppm and 2 000 ppm.
While the two outputs always have the same reporting range, the span, signal offset and scaling of their
electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 6.13.4
for instructions on adjusting these parameters.
To select SNGL range mode and to set the upper limit of the range, press:
SETUP C.3
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP C.3 RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
EXIT x 2 returns
to the main
SAMPLE display
SETUP C.3 RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP C.3 RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP C.3 RANGE CONTROL MENU
MODE
SET UNIT EXIT
SAMPLE* RANGE = 500.000 PPM CO2 X.XXX
< TST TST > CAL SETUP