To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (See Section
6.13.1), then press:
DIAG AIO AUTO CALIBRATING CONC_OUT_1
AUT O CALIBRATING CONC_OUT_2
AUTO CALIBRATING TEST_OUTPUT
Exit at any time
to return to the
main DIAG
menu
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO
A
OUTS CALIBRATED: NO
< SET SET> CAL EXIT
If AutoCal has been
turned off for any
channel, the message
for that channel will be
similar to:
NOT AUTO CAL
CONC_OUT_1
Exit to return to
the I/O
configuration
menu
STARTING FROM DIAGNOSTIC MENU
(see Section 6.13.1)
If any of the channels have
not been calibrated this
message will read NO.
DIAG AIO AOUTS CALIBRATED: YES
< SET SET> CAL EXIT
To automatically calibrate a single analog channel, activate the ANALOG I/O CONFIGURATION MENU
(See Section 6.13.1), then press: