Teledyne gfc 7000ta Carbon Monoxide Alarm User Manual


 
Theory of Operation Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 191
IR shining through both cells is effected
equally by interfering gas in the Sample
Chamber
M/R
is Shifted
Figure 9-5: Effects of Interfering Gas on CO2 MEAS & CO2 REF
Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO
2
and not to
interfering gases. In this way, Gas filter correlation rejects the effects of interfering gases and so that the
analyzer responds only to the presence of CO
2
.
To improve the signal-to-noise performance of the IR photo-detector, the GFC wheel also incorporates an
optical mask that chops the IR beam into alternating pulses of light and dark at six times the frequency of
the measure/reference signal. This limits the detection bandwidth helping to reject interfering signals
from outside this bandwidth improving the signal to noise ratio.
The IR Signal as the Photo-Detector sees it
after being chopped by the GFC Wheel
Screen
CO
2
REF
CO
2
MEAS
Figure 9-6: Chopped IR Signal