Teledyne gfc 7000ta Carbon Monoxide Alarm User Manual


 
Operating Instructions Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments 66
To view the TEST Functions press:
RANGE
RANGE1
1
RANGE2
1
O
2
RANGE
STABIL
MEAS
REF
MR RATIO
PRES
SAMP FL
SAMP TEMP
BENCH TEMP
WHEEL TEMP
BOX TEMP
PHT DRIVE
SLOPE
OFFSET
TEST
TIME
SAMPLE RANGE = 500.000 PPM CO2 = XXX.X
< TST TST > CAL SETUP
1
Only appears instrument is set
for DUAL or AUTO reporting
range modes.
2
Only appears if O
2
Sensor
Option is installed.
Refer to
Table 6-2 for
definitions of
these test
functions.
Toggle <TST TST> buttons
to scroll throu
g
h list of
Figure 6-2: Viewing TEST Functions
NOTE
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-
range reading or the analyzer’s inability to calculate it.
All pressure measurements are represented in terms of absolute pressure.
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1
in-Hg per 300 m gain in altitude. A variety of factors such as air conditioning and
passing storms can cause changes in the absolute atmospheric pressure.