Emerson PN 51-1055pHC/rev.K Water System User Manual


 
MODEL SOLU COMP II SECTION 8.0
TROUBLESHOOTING
FIGURE 8-7. Checking for a Poisoned Reference
Electrode.
Refer to the sensor wiring diagram to identify the reference leads.
A laboratory silver/silver chloride electrode can be used in place
of the second sensor.
75
8.9 MEASURING REFERENCE VOLTAGE
Some processes contain substances that poison or
shift the potential of the reference electrode. Sulfide
is a good example. Prolonged exposure to sulfide
converts the reference electrode from a silver/silver
chloride electrode to a silver/silver sulfide electrode.
The change in reference voltage is several hundred
millivolts. A good way to check for poisoning is to
compare the voltage of the reference electrode with
a silver/silver chloride electrode known to be good.
The reference electrode from a new sensor is best.
See Figure 8-7. If the reference electrode is good,
the voltage difference should be no more than
about 20 mV. A poisoned reference electrode usu-
ally requires replacement.