Emerson PN 51-1055pHC/rev.K Water System User Manual


 
58
MODEL SOLU COMP II SECTION 6.0
CALIBRATION
a. Press MENU. The main menu screen appears. Choose Program.
b. Choose Measurement.
c. Choose Sensor2 (conductivity).
d. Choose Cond.
e. Choose Other.
f. Choose Raw.
g. The display returns to the screen in step c. Return to the main display
by pressing MENU followed by EXIT.
a. Press MENU. The main menu screen appears. Choose Calibrate.
b. Choose Sensor2 (conductivity).
c. Choose Measurement.
d. Choose InProcess.
e. The first line is the current (or live) conductivity reading. Change the
number in the second line to match the conductivity of the standard.
Press ENTER. Be sure the live reading is stable before making
changes.
f. If the calibration is acceptable, the updated cell constant screen
appears for 5 seconds. The display returns to the screen in step b.
g. If the calibration is unacceptable (cell constant is less than 0.0001/cm
or greater than 100/cm), the Calibration Error screen appears for 5
seconds. The display then returns to the screen in step e. Repeat the
calibration.
Calibrate Hold
PPrrooggrraamm
Display
CCaalliibbrraattee
Hold
Program Display
Configure?
SSeennssoorr11
Sensor2
Calibrate?
SSeennssoorr11
Sensor2
Cal Sensor2?
MMeeaassuurreemmeenntt
Temp
Outputs Alarms
MMeeaassuurreemmeenntt
>>
S2 Temp Corr?
NNeeuuttSSaalltt
Other
S2 Temp Corr?
Slope Cation
RRaaww
S2 Measure?
CCoonndd
TDS Resistivity
CalS2?
IInnPPrroocceessss
Meter CellConst
Live 10.00µS/cm
CalS2
11
0.00µS/cm
Updated S2 Cell
Const: 1.0013/cm
Calibration
Error
6. Make the Solu Comp II reading agree with the known conductivity.