5. Choose the desired language. Select >> to show more choices.
6. Choose the number of sensors wired to the analyzer.
NOTE
If One sensor is chosen, only
S1 (pH/ORP) will be available
.
S2 (conductivity) cannot be chosen for single measurement.
7. Select the measurement for sensor 1.
8. Select the measurement for sensor 2. This screen will not appear if one sensor
is selected.
9. Enter the cell constant for the conductivity sensor (sensor 2). See label attached
to sensor.
10. Choose temperature units.
11. The main display appears. The outputs and alarms are assigned to default val-
ues. To change settings, refer to Section 5.0, Programming the Analyzer. To
reinstall factory settings and return to Quick Start, see Section 5.10.
# of sensors?
One
TTwwoo
Temperature in?
CC
F
S1 Measure?
ppHH
Redox ORP
S2 Measure?
CCoonndd
TDS Resistivity
4. When the analyzer is powered up for the first time, Quick Start screens appear. Using Quick Start is easy.
a. A blinking field shows the position of the cursor.
b. Use the or key to move the cursor left or right. Use the or key to move the cursor up or down or to
increase or decrease the value of a digit. Use the or key to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the previous screen.
EEnngglliisshh
Fran ais
Espa ol >>
&
If there is no cell constant on the label, calculate it
from the equation:
cell const = K
500 + cal const
1000
e
j
Cell Constant?
S1:
11
.0000/cm