Emerson PN 51-1055pHC/rev.K Water System User Manual


 
MODEL SOLU COMP II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES AND APPLICATIONS
1.2 SPECIFICATIONS
1.3 ORDERING INFORMATION AND ACCESSORIES
1
1.1 FEATURES AND APPLICATIONS
The Solu Comp II analyzers offer the choice of sin-
gle or dual sensor input with measurement choices
of pH/ORP, resistivity/conductivity/TDS, % concen-
tration, ratio conductivity, total and free chlorine, dis-
solved oxygen, dissolved ozone, flow and tempera-
ture. Dual measurement analyzers offer a wide
choice of measurement combinations thus reducing
the cost per loop and needed panel space.
FIELD COMMISSION OPTION: The Solu Comp II
can be ordered with the ability to commission meas-
urements in the field. This added flexibility can
greatly reduce the number of spare instruments
required for field servicing.
QUICK START PROGRAMMING: Exclusive Quick
Start screens appear the first time the Solu Comp II
is powered up. Screen prompts direct the user to
register the number of sensors, the measurement
unit(s) and the language to display. Some measure-
ment specific prompts are also displayed. The
measurement loop is ready for use in a matter of
minutes.
MENUS: Menu screens for calibrating and register-
ing choices are simple and intuitive. Plain language
prompts guide the user through the procedures.
There are no service codes to enter before gaining
access to menus.
DUAL SENSOR INPUT AND OUTPUT: The Solu
Comp II accepts single or dual sensor input. The
two 4-20 mA outputs can be independently pro-
grammed to correspond to any selected measure-
ment or temperature. Output damping and linear or
log output may also be field selected.
ALARMS: The Solu Comp II has three fully pro-
grammable alarm relays that can be assigned to
any selected measurement or temperature. Alarms
can be configured as high, low, or USP
1
. The third
relay has the additional choice of fault alarm opera-
tion. When selected, a fault alarm will activate the
relay when a sensor or analyzer fault occurs.
ENCLOSURE: The panel mount version fits stan-
dard ½ DIN panel cutouts, and its shallow depth is
ideally suited for easy mounting in Hoffman-type
enclosures. A panel mount gasket is included to
maintain the weather rating of the panel.
Surface/pipe mount enclosure includes self-tapping
screws for surface mounting. A pipe mounting
accessory kit is available for mounting to a 2-inch
pipe.
DISPLAY: The two-line, 16-character, back-lit dis-
play can be customized to meet user requirements.
All operations and descriptive messages can be
field selected for English, French, German, Italian,
Spanish, or Portuguese. Informative screens, which
permit data not shown in the regular display, may
be seen at the push of a button.
TEMPERATURE: Most measurements (except
ORP and flow) require temperature compensation.
The Solu Comp II will automatically recognize either
a Pt100 or Pt1000 RTD, normally built into the sen-
sor. When this RTD is present, the Solu Comp II
can be set up to display the temperature in °C or °F
as well as set any one or more of the alarms and/or
outputs to respond to this sensor input. If two meas-
urements with temperature are present either can
be chosen for each alarm and output selected.
1
USP alarm applies to conductivity/resistivity only.