36
MODEL SOLU COMP II SECTION 5.0
PROGRAMMING THE ANALYZER
5.4.4 Procedure: Programming Alarm Setpoints
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Alarms.
3. Choose Alarm Setpoints.
4. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).
5. The display shows the alarm selected (AL1) and the configuration. The
alarm is for Sensor 1 (S1), and the logic is high. Use the arrow keys to
change the alarm setpoint.
6. If the alarm is USP 24, the display shows the alarm selected and the USP
safety margin. Change the safety margin to the desired value.
7. The display returns to the Select Alarm? screen. Select another alarm or
press EXIT to return to the previous screen. To return to the main display,
press MENU followed by EXIT.
Calibrate Hold
PPrrooggrraamm
Display
AAllaarrmm SSeettppooiinnttss
Alarm Configure
Select Alarm?
AALL11
AL2 AL3
AL1 S2 Setpoint?
USP24
11
0% Safety
AL1 S1 Setpoint?
High
11
4.00pH
Outputs
AAllaarrmmss
Measurement >>