Emerson PN 51-1055pHC/rev.K Water System User Manual


 
35
MODEL SOLU COMP II SECTION 5.0
PROGRAMMING THE ANALYZER
5.4.3 Procedure: Configuring Alarms
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Alarms.
3. Choose Alarm Configure.
4. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).
5. For AL1 or AL2
a. Choose Sensor 1 (pH) or Sensor 2 (conductivity).
b. Choose Measurement or Temp.
c. Choose High, Low or USP 24. USP 24 appears only if Sensor 2 was
selected.
d. Set the alarm Deadband.
6. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
7. For AL3
a. Choose Sensor1 (pH), Sensor2 (conductivity), or Fault.
b. For sensor 1 or 2, choose Measurement or Temp.
c. Choose High, Low or USP 24. USP 24 appears only if Sensor 2 was
selected. Set the deadband.
d. Choosing Fault means AL3 will activate when a sensor or analyzer
fault exists. There is no user setting to make.
8. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.