MODEL SOLU COMP II SECTION 5.0
PROGRAMMING THE ANALYZER
5.5.4 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Measurement.
3. Choose Sensor 1 (pH) or Sensor 2 (conductivity). For a single input con-
figuration, the Sensor 1 Sensor 2 screen does not appear.
If you chose Sensor 1, go to step 4.
If you chose Sensor 2, go to step 11.
4. For Sensor 1, choose pH, Redox, or ORP.
If you chose pH, do steps 5 through 9.
If you chose ORP or Redox, do step 10.
5. Choose NO if the pH sensing electrode is NOT a glass electrode.
6. Choose Soln Temp Corr or Sensor Isoptntl.
7. For Soln Temp Corr, choose Off, UltraPure, HighpH, or Custom. For
Custom, enter the desired temperature coefficient.
8. For Sensor Isoptntl, enter the desired sensor isopotential pH. Do not
change the sensor isopotential pH unless the sensor is known to have an
isopotential pH different from 7.00.
9. The display returns to the screen shown in step 3. Choose Sensor 2
(conductivity) and follow the prompts. See step 11.
10. If Redox or ORP was selected, there are no further settings to make. The
display returns to the screen shown in step 3. Choose Sensor 2 (con-
ductivity) and follow the prompts. See step 11.
11. For Sensor 2, choose Cond (conductivity), Resistivity, or TDS (total dis-
solved solids).
If you chose Cond or Resistivity, go to step 12.
If you chose TDS, go to step 14.
12. Choose NeutSalt, Slope, Cation, or Raw. For Slope, enter the desired
temperature coefficient in %/°C.
13. The display returns to the screen shown in step 3. Press EXIT to return to
the previous screen. To return to the main display, press MENU followed
by EXIT.
14. If TDS was selected, there are no further settings to make.The display
returns to the screen shown in step 3. Press EXIT to return to the previ-
ous screen. To return to the main display, press MENU followed by EXIT.
Calibrate Hold
PPrrooggrraamm
Display
Config?
SSeennssoorr11
Sensor2
S2 Temp Corr?
NNeeuuttSSaalltt
Other
S1Measure?
ppHH
Redox ORP
S2 Measure?
CCoonndd
TDS Resistivity
S1 SolnTempCorr?
OOffff
Ultrapure >>
Sensor Isoptntl
S1:
00
7.00pH
S1 Glass Fault
Enable?
YYeess
No
Outputs Alarms
MMeeaassuurreemmeenntt
>>
SSoollnn TTeemmpp CCoorrrr
Sensor Isoptntl
39