Emerson PN 51-1055pHC/rev.K Water System User Manual


 
MODEL SOLU COMP II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
2
1.2 SPECIFICATIONS - General
Case: ABS (panel-mount), polycarbonate (pipe- and
surface-mount). All versions are NEMA 4X/CSA 4
(IP65).
Dimensions
Panel (code -10): 6.10 x 6.10 x 3.72 in. (155 x
155 x 94.5 mm)
Surface/Pipe (code -11): 6.23 x 6.23 x 3.23 in.
(158 x 158 x 82 mm); see page 5 for dimensions
of pipe mounting bracket.
Conduit openings: Accepts PG13.5 or 1/2 in. con-
duit fittings
Display: Two line, 16-character, back-lit display.
Character height: 4.8 mm. Display can be cus-
tomized to meet individual requirements.
Depending on number of sensors, as many as 14
display screens are available.
Ambient temperature and humidity: 0 to 50°C, (32
to 122°F) RH 5 to 95% (non-condensing)
Note: The analyzer is operable from -20 to 60°C
(-4 to 140°F) with some degradation in display
performance.
Power:
Code -01: 115/230 Vac ±15%, 50/60 Hz ±6%, 8.0W
Code -02*: 24 Vdc ±15%, 6.0W
Installation Category II
*
For +24Vdc Power Supply use only devices meeting
NEC Class II or UL recognized (UL 1950).
Equipment protected throughout by double insula-
tion.
Hazardous Location:
Class I, Division 2,
Groups A, B, C, & D
POLLUTION DEGREE 4: Extended Environment
Outdoor use where conductive contamination
such as rain, snow, or dust may be present.
(Hazardous Location only)
RFI/EMI: EN-61326
LVD: EN-61010-1
Input: Choice of single or dual sensor input with
measurement choices of pH/ORP, conductivity/
resistivity, toroidal conductivity, flow, chlorine, dis-
solved oxygen, and dissolved ozone. Field-com-
missioned units allow user to change measure-
ments on either or both inputs. See combination
guide for valid combinations. For contacting con-
ductivity measurements, temperature element
must be a Pt 1000 RTD. For other measure-
ments, use either a Pt100 RTD, Pt1000 RTD, or
22k NTC (D.O. only).
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Linear or logarithmic.
Maximum load 500 ohms. Output dampening with
time constant of 5 sec is user-selectable.
Alarms: Three alarm relays for process measure-
ment(s) or temperature. Alarm 3 can be
configured as a fault alarm, instead of a
process alarm. Each relay can be configured
independently. Alarm logic (high or low activation
or USP*) and deadband are user-programmable.
The USP* alarm can be programmed to activate
when the conductivity is within a user-selectable
percentage of the limit.
*conductivity/resistivity measurement only
Relays: Form C, single pole double throw, epoxy sealed
Terminal Connections Rating: 26-14 AWG wire size
Weight/Shipping weight (rounded up to nearest lb or
nearest 0.5 kg): 3 lb (1.5 kg)/4 lb (2.0 kg)
Resistive Inductive
28 Vdc 5.0 A 3.0 A
115 Vac 5.0 A 3.0 A
230 Vac 5.0 A 1.5 A
Ordinary Location: (-68 only)
POLLUTION DEGREE 2: Normally only non-con-
ductive pollution occurs. Occasionally, however, a
temporary conductivity caused by condensation
must be expected.
12RN