b. The equipment complies with relevant IEC and ISO safety standards and
is supplied from a medical grade isolation transformer.
c. The equipment complies with relevant IEC and ISO safety standards and
is kept at least 1.5 meters from the patient.
The allowable leakage currents of IEC 60601-1-1 must not be exceeded. IEC
60601-1-1 should be consulted when assembling such a system.
Electromagnetic compatibility
The FP35 complies with IEC 60601-1-2.
The FP35 generates and uses radio frequency energy. In some cases the FP35
could cause interference to radio or television reception. You can determine if
the FP35 is the source of such interference by turning the unit off and on.
If you are experiencing interference caused by the FP35, you may be able to cor-
rect it by one or more of the following measures:
1. Relocate or reorient the receiving antenna.
2. Increase the distance between the FP35 and the receiver.
3. Connect the FP35 to a different outlet than the receiver.
In some cases radio transmitting devices, such as cellular telephones, may cause
interference to the FP35. In this case try increasing the distance between the
transmitter and the FP35.
Disposal of the FP35 and accessories
The FP35 and some of its accessories contain lead. At the end of its useful life,
please recycle or dispose of the FP35 according to local regulations.
If you are located in the European Union, please report all safety-related con-
cerns to our authorized representative:
Siemens Hearing Instruments Ltd.
Alexandra House
Newton Road
Manor Royal
Crawley
West Sussex RH109TT
ENGLAND
Otherwise, please report all safety related concerns to the Frye factory:
Frye Electronics, Inc.
9826 SW Tigard St.
Tigard, OR 97223
Ph: (503) 620-2722 or (800) 547-8209
Fax: (503) 639-0128
email: support @frye.com
sales@frye.com
An Overview of the FP35
15