152 FONIX FP35 Hearing Aid Analyzer
Figure 5.7.3.2—Real-Ear IM distortion sweep
13. Set up the client, the aid, and the analyzer for real-ear testing. Make sure to
level the sound field speaker.
14. Press [START/STOP] to start the IM distortion sweep. See Figure 5.7.3.2.
The measurement will stop when finished. The percentage scaling for the
measurement is located on the right side of the graph. Notice the thick line
below the lower right corner of the graph. This is an indication that the cur-
rent selected measurement uses the right side scaling.
5.7.3.3 To change the frequencies used in the sweep
By default, the IM distortion sweep is performed at the frequencies used in the
NORMAL pure-tone sweep. This can be changed to perform the measurements
at fewer frequencies (for a faster test), and more frequencies (for a more com-
plete test).
1. Press [MENU] from the real-ear measurement screen to enter the real-ear
menu.
2. Press [NEXT] twice to enter the Custom Menu.
3. Use [
∨, ∧] to select DISTORTION SWEEP.
4. Use [<, >] to make one of the following selections:
• NORM:Tomeasureat1/12octavefrequenciesclosestto100Hzintervals
(default test)
• FAST:Tomeasureat1/3octavefrequenciesclosestto100Hzintervals
(fastest test)
• FULL:Tomeasureatevery100Hzinterval(mostcompletetest)