A SERVICE OF

logo

114 FONIX FP35 Hearing Aid Analyzer
• Makesureallcablesinthecompartmentareputawaysotheywillnot
interfere with replacing the speaker.
• Setthespeakerbackinthecompartment.Asyoudothis,guidethespeak-
er wire so it stays to the side of the speaker, not underneath it.
5.2.3 External Speaker Setup
If desired, you can use an external speaker to perform your real-ear measure-
ments. The external speaker can be mounted to a floor stand or to a swing arm.
A swing arm is especially nice because you can move the speaker around the
patient when you switch ears, rather than move your patient around the speak-
er.
Figure 5.2.3—External sound field speaker setup
To set up your FP35 software to use an external speaker:
1. Press [MENU] from the Opening screen to enter the Default Settings menu.
2. Press [NEXT] to enter the Advanced Default Settings menu.
3. Select SPEAKER under RESETS with [
∨, ∧] keys.
4. Set AUTO with [<, >]. All real-ear measurements will now be done with
the external speaker and all coupler measurements will be done with the
internal speaker.
5. Press [EXIT] to return to the Opening screen.
6. Press [F2] to enter Real-Ear Mode again.