3 Operation Model 6600
3-16 Part I
Teledyne Analytical Instruments
-Enter the System Menu
-Select MORE on the first and second System menu screens.
-In the last System Menu screen you will see:
ALGORITHM APPLCATION
MODEL OUT_CAL ANLZ
-Select the last field “ANLZ” using the Up/Down switch.
-Press Enter to change the mode of the filter and the solenoid. The se-
quence is as follows:
1. ZERO: Sets the Filter and solenoid in the zero mode (span filter off,
zero solenoid on).
2. SPN1: Sets the Filter and solenoid in the span mode (span filter on,
zero solenoid on).
3. SPN2: Sets second span filter on (usually not installed) and zero
solenoid on
4. SPNB: Sets both span filter on (usually only one filter installed) and
zero solenoid on.
5. ANLZ: Returns Filters and solenoid to the Analyze mode (span filter
off, zero solenoid off).
-Press Escape to see the effect
NOTE: FOR PROPER OPERATION OF THE ANALYZER
RETURN TO ANLZ MODE.
3.4 The Zero and Span Functions
The Model 6600 can have as many as three analysis ranges plus a special
calibration range (Cal Range). Calibrating any one of the ranges will automatical-
ly calibrate the other ranges.
CAUTION: Always allow one hour warm-up time before calibrat-
ing, if your analyzer has been disconnected from its
power source. This does not apply if the analyzer
was plugged in but was in STANDBY.
The analyzer is calibrated using zero, and span gases.