Teledyne 6600 Water System User Manual


 
Oil in Water Analyzer Operation 3
Part I 3-1
Teledyne Analytical Instruments
Operation
3.1 Introduction
Although the Model 6600 is usually programmed to your application at the
factory, it can be further configured at the operator level, or even, cautiously,
reprogrammed. Depending on the specifics of the application, this might include
all or a subset of the following procedures:
Setting system parameters:
Establish a security password, if desired, requiring Operator
to log in (secure in safe file for referrence).
Establish and start an automatic calibration cycle, if desired.
Routine Operation:
Calibrate the instrument.
Choose autoranging or select a fixed range of analysis.
Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
Program/Reprogram the analyzer:
Define new applications.
Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
3.2 Using the Controls
To get the proper response from these controls, turn the control toward the
desired action (ESCAPE or ENTER—DOWN or UP), and then release it.
Turn-and-release once for each action. For example, turn-and-release twice
toward UP to move the VFD screen two selections upwards on the list of
options (menu).