3 Operation Model 6600
3-12 Part I
Teledyne Analytical Instruments
1 4.5
2 5.0
3 5.0
4 5.5
5 7.0
6 9.0
7 14.0
8 25.0
9 46.0
10 90.0
At a setting of “zero”, the response time is purely set by the electronics to
4.5 seconds. The numbers above can and will change depending on application
and they merely serve to illustrate the effect of the digital filter.
3.3.8 Homogenizer Function Setup
Depending on the application, the 6020 sampling system may have an
Ultra Sonic Homogenizer. The function of this part is to prevent the oil in the
water from clumping together. The homogenizer should turn on after the initial
warm up and self-diagnostic period when the analyzer enters the Analyze mode.
The homogenizer will turn off automatically as soon as the analyzer enters the
Zero and Span mode, turning back on at the end.
Under some conditions, it might be desirable to manually turn the Ultra
Sonic Homogenizer off. The homogenizer set up function is provided in the
System menu. To access the Homogenizer function setup:
1. Enter the System menu.
2. Select MORE in the first System menu using the UP/DOWN switch.
3. Select HMGNZR in the second System menu, the screen will display
Set Ultra Sonic
Homogenizer: ON
4. By using the UP/DOWN switch the homogenizer can be toggled on and off.
5. Enter or Escape to return to the analyze mode.