Teledyne 6600 Water System User Manual


 
A-2
Appendix Models 6600
Teledyne Analytical Instruments
Typical Analytical Performance Specifications:
(
will vary per application)
Accuracy: ±2% of full scale possible (Oil in Water)
When calibrated on specific oil of interest.
Noise: Less than ±1%
Drift: Less than ±1% per auto Zero cycle (source/
detector dependent)
Sample Cell: (Aluminum/CPVC for Oil in Water) with
Sapphire/Quartz windows standard. (Kynar,
Kalrez optional)
Cell Length: 1/8 to 3.5” (application dependent)
Flow Rate: 100 ccm to 2 lpm (application dependent)
Light Source: Mercury, Hg
Sensitivity: .08 to 1.0 absorbance units.
Reproducibility: +/-1% of scale or better
Filter Wavelength: 210 to 1000 millimicrons. (application depen-
dant - oil in water 254/365mm).
Sample Pressure: 150 psi maximum (oil in water applications)
Sample temperature: 1-120
o
C (34-250
o
F) (non freezing, non-boiling)