Fluke 5020A Thermostat User Manual


 
Digital Communications Interface
Commands 6
6-27
DATa:DSTatistics:RHOur? [MIN|MAX|DEF]
Returns the daily statistics reset time.
Example: DAT:DST:RHO?
Response: 12
Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or
default value respectively.
DATa:DSTatistics:RHOur <num>|MIN|MAX|DEF
Sets the daily statistics reset time.
Example: DAT:DST:RHO 0
The <num> parameter specifies the hour for the daily statistics reset time. Specifying
MIN, MAX, or DEF parameter sets the daily statistics reset time to the minimum,
maximum, or default value respectively. This command may be password protected (see
Password Commands).
DATa:DSTatistics:RECord:BTIMe? [<num>]
Returns the reset time for a daily statistics set.
Example: DAT:DST:REC:BTIM?
Response: 0,0,0
The <num> parameter specifies the daily statistics record number. If none is specified
then record 1 is assumed. The response is returned in the format,
<hour>,<minute>,<second>.
DATa:DSTatistics:RECord:CLEar
Clears all daily statistics records.
Example: DAT:DST:REC:CLE
This command may be password protected (see Password Commands).
DATa:DSTatistics:RECord:COUNt? [<MAX>]
Returns the daily statistics record count.
Example: DAT:DST:REC:COUN?
Response: 64
Specifying the MAX parameter returns the maximum number of daily statistics records
than can be held in memory.
DATa:DSTatistics:RECord:DATE? [<num>]
Returns the daily statistics recording date.
Example: DAT:DST:REC:DATE?
Response: 2003,9,12
The <num> parameter specifies the daily statistics record number. If none is specified,
record 1 is assumed. The response is returned in the format, <year>,<month>,<day>.