Fluke 5020A Thermostat User Manual


 
5020A
Users Manual
6-16
Response: 20.00
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). Specifying the MIN, MAX,
or DEF parameter returns the minimum, maximum, or default value respectively.
ALARm:RHUMidity<chn>:LOWer:LIMit <float>|MIN|MAX|DEF
Sets the lower humidity alarm limit for the specified channel in %RH.
Example: ALAR:RHUM1:LOW:LIM 25.00
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <float> parameter
specifies the value for the lower humidity alarm limit. Specifying the MIN, MAX, or
DEF parameter sets the lower humidity alarm limit to the minimum, maximum, or default
value respectively. This command may be password protected (see Password
Commands).
ALARm:RHUMidity<chn>:RATE?
Indicates whether a humidity rate alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:RATE?
Response: 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if a
humidity rate alarm event has occurred on the specified channel. A value of 0 is returned
if a humidity rate alarm event has not occurred on the specified channel.
ALARm:RHUMidity<chn>:RATE:ENABle?
Returns the state of the humidity rate alarm enable for the specified channel.
Example: ALAR:RHUM1:RATE:ENAB?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if
the humidity rate alarm is enabled. A value of 0 is returned if the humidity rate alarm is
disabled.
ALARm:RHUMidity<chn>:RATE:ENABle <bool>
Sets the state of the humidity rate alarm enable for the specified channel.
Example: ALAR:RHUM1:RATE:ENAB 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <bool> parameter
turns the humidity rate alarm enable on (1 or ON) or off (0 or OFF). This command may
be password protected (see Password Commands).
ALARm:RHUMidity<chn>:RATE:LIMit? [MIN|MAX|DEF]
Returns the humidity rate alarm limit for the specified channel in %RH per hour.
Example: ALAR:RHUM1:RATE:LIM?
Response: 5
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). Specifying the MIN, MAX,
or DEF parameter returns the minimum, maximum, or default value respectively.