Fluke 5020A Thermostat User Manual


 
Digital Communications Interface
Commands 6
6-17
ALARm:RHUMidity<chn>:RATE:LIMit <float>|MIN|MAX|DEF
Sets the humidity rate alarm limit for the specified channel in %RH per hour.
Example: ALAR:RHUM1:RATE:LIM 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <float> parameter
specifies the value for the humidity rate alarm limit. Specifying the MIN, MAX, or DEF
parameter sets the humidity rate alarm limit to the minimum, maximum, or default value
respectively. This command may be password protected (see Password Commands).
ALARm:RHUMidity<chn>:SENSor?
Indicates whether a humidity sensor alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:SENS?
Response: 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if a
humidity sensor alarm event has occurred on the specified channel. A value of 0 is
returned if a humidity sensor alarm event has not occurred on the specified channel.
ALARm:RHUMidity<chn>:SENSor:ENABle?
Returns the state of the humidity sensor alarm enable for the specified channel.
Example: ALAR:RHUM1:SENS:ENAB?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if
the humidity sensor alarm is enabled. A value of 0 is returned if the humidity sensor
alarm is disabled.
ALARm:RHUMidity<chn>:SENSor:ENABle <bool>
Sets the state of the humidity sensor alarm for the specified channel.
Example: ALAR:RHUM1:SENS:ENAB 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <bool> parameter
turns the humidity sensor alarm on (1 or ON) or off (0 or OFF). This command may be
password protected (see Password Commands).
ALARm:RHUMidity<chn>:UPPer?
Indicates whether an upper humidity alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:UPP?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if
an upper humidity alarm event has occurred on the specified channel. A value of 0 is
returned if an upper humidity alarm event has not occurred on the specified channel.
ALARm:RHUMidity<chn>:UPPer:ENABle?
Returns the state of the upper humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:UPP:ENAB?
Response: 0