Fluke 5020A Thermostat User Manual


 
5020A
Users Manual
6-50
measure bit.
Bit 3 Channel 2 humidity measure
bit.
Bit 4 Not used
Bit 5 Not used
Bit 6 Not used.
Bit 7 Not used.
Example: STAT:MEAS:COND?
Response: 15
STATus:MEASure:ENABle?
This command returns the Measurement Status Enable Register (see
STATus:MEASure:ENABle <num> command below).
Example: STAT:MEAS:ENAB?
Response: 15
STATus:MEASure:ENABle <num>|MIN|MAX|DEF
This command sets the Measure Status Enable Register. This register determines which
event bits of the Measure Status Event Register affect the Measure Status Bit of the
Status Byte Register. If any bit in the Measure Status Event Register is set (1) while its
corresponding mask bit in the Measure Status Enable Register is set, the Measure Status
Bit in the Status Byte Register will be set. The value of the <num> parameter is a number
from 0 to 15 that is the sum of the binary-weighted values of each mask bit. The Measure
Status Event Register is described in Section 8.4.11.13, STATus:MEASure?. The
following example causes the Measure Status Bit in the Status Byte Register to be set
when measurements on both channels have been completed.
Example: STAT:MEAS:ENAB 15
Response: 15
STATus:OPERation?
This command reads and clears the Operation Status Event Register. The value returned
indicates the condition of each of the eight bits of the register by adding the binary-
weighted values of each bit. The meaning of each bit, when set (1), is as follows where 0
is the least significant and 7 is the most significant:
0: no measurement
Bit 0 Not used.
Bit 1 Not used.
Bit 2 Not used.
Bit 3 Not used.
Bit 4 Measurement Complete bit.
Bit 5 Not used.
Bit 6 Not used.
Bit 7 Not used.
Example: STAT:OPER?
Response: 16
STATus:OPERation:CONDition?
This command reads the Operation Status Condition Register. Reading this register does
not affect it. The value returned indicates the condition of each of the eight bits of the