Data Record
The DATA RECORD function allows the setting of the data recording features. The
functions that appear in this menu are RECORD SETTING, DATA VIEW, DATA
PRINT, DATA STORAGE, and DATA CLEAR.
avu018.jpg
Figure 5-17. Data Record
Use the A D buttons to select the menu. Press the E button to select the menu. Press
the F button to cancel any changes and exit to the menu.
Record Setting
The RECORD SETTING function allows you to select what is to be recorded and how
often the recording occurs. This may be password-protected. The temperature and
humidity from each channel can be recorded and the recording period is selectable.
To record the temperature on sensor channel 1, set REC 1/T to ON.
To record the humidity on sensor channel 1, set REC 1/H to ON.
To record the temperature on sensor channel 2, set REC 2/T to ON.
To record the humidity on sensor channel 2, set REC 2/H to ON.
The REC PER recording period is set by selecting 1, 2, 5, 10, 15, or 30 seconds, 1, 2, 5,
10, 20, or 30 minutes, or 1 hour for the REC PER.
Note
It is recommended that REC PER be set to 5 minutes or more; otherwise, the memory may fill up
quickly and redrawing graphs may be slow. Also note that graphs can only be redrawn if the
measurement for the channel is being recorded, as the graph data is obtained from recorded
data. The recording period affects the observable time resolution of graphs
.