Table 7-3. Agilent SAS Programming Parameters (in Fixed Mode)
Parameter Agilent E4350B Agilent E4351B
CURR[:LEV] MAX and
8.16 A 4.08 A
CURR[:LEV]:TRIG MAX
8.16 A 4.08 A
CURR:PROT[:LEV]: MAX
10 A 5 A
(Programming range is 0 to MAX)
*RST Current Value
0.096 A 0.048 A
*RST Current Protect Value
MAX for both models
OUTP:PROT:DEL
*RST Value
0 to 32.767 s (MAX) both models
200 ms both models
VOLT[:LEV] MAX and
61.50 V 123.0 V
VOLT[:LEV]:TRIG MAX
61.50 V 123.0 V
(Programming range is 0 to MAX)
*RST Voltage Value
0 V for both models
VOLT:PROT MAX
73.0 V 140.0 V
(Programming range is 0 to MAX)
*RST OVP value
73.0 V 140.0 V
Calibration Commands
See appendix B.
Display Subsystem
This subsystem controls the state and output of the alphanumeric portion of the display.
DISP
Enables or disables the display. When disabled, the display characters are blank. The annunciators are not affected by this
command.
Command Syntax DISPlay[:WINDow][:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST Value ON
Examples DISP ON DISPLAY:STATE ON
Query Syntax DISPlay[:WINDow][STATe]?
Returned Parameters <NR1> 0 or 1
Related Commands DISP:MODE DISP:TEXT *RST
DISP:MODE
Switches the display between its normal metering mode and a mode in which it displays text sent by the user.
Command Syntax DISPlay[:WINDow]:MODE <CRD>
Parameters NORMal | TEXT
*RST Value NORM
Examples DISP:MODE NORM DISPLAY:MODE NORMAL
Query Syntax DISPlay[:WINDow]:MODE?
Returned Parameters <CRD> NORMAL or TEXT
Related Commands DISP DISP:TEXT *RST