Agilent Technologies E4351B Water Heater User Manual


 
User Connections 39
Note The Agilent SAS is shipped from the factory with its GPIB address set to 5. The Agilent SAS primary
and secondary addresses can be changed from the front panel as described under Changing the GPIB
Address in chapter 5. For Agilent SAS GPIB interface capabilities, see appendix A.
ô
í
÷
û
ø
ù
From 1 to 15 direct supplies may be connected to 1 controller GPIB interface.
Tighten connector thumbscrews by hand. Do not use a screwdriver.
Do not stack more than 3 connectors on a GPIB receptacle.
GPIB cable (see Accessories in Chapter 1)
From 1 to 15 linked supplies may be connected to 1 direct unit.
Either receptacle (Jl or J2) may be used as an input or an output.
Serial Link Cable (see Accessories in Chapter 1), 2 meters. 1 is supplied.
Maximum total length of all GPIB cables (including controller) not to exceed 20 meters.
Use caution with individual lengths over 4 meters.
Maximum total length of all serial cables not to exceed 30 meters.
1.
2.
3.
NOTES:
A direct Agilent SAS is connected to the controller interface and must have a unique primary GPIB bus address.
The stand-alone configuration uses only direct supplies connected to the controller interface.
The linked configuration uses 1 or more linked units connected to each direct unit. Each linked unit has a unique
secondary GPIB bus address and derives its primary address from the direct unit.
Figure 4-12. Controller Connections