Teledyne 300 Carbon Monoxide Alarm User Manual


 
If none of these help, see the TROUBLESHOOTING SECTION 10.0.
Record all the Analyzer setup data from the display.
Range
DCPS
Box Temp
Wheel Temp
Bench Temp
Sample Temp
Sample Flow
Pressure
MR ratio (Measure/reference ratio)
CO Reference
CO Measure
These data can be useful in future troubleshooting.
8.3 ZERO/SPAN CHECKING
It is recommended that the Model 300 be checked daily for zero and span drift.
The EPA requires that the Analyzer be zero and span checked at least once every two
weeks.
With the Automatic Zero/Span Check and Remote Zero/Span Check features of the
Model 300, daily zero and span checks are easy.
For Automatic Zero/Span Check and Remote Zero/Span Check to be effective, it is
necessary that the Analyzer have the Zero/Span valve option. This option includes the
two three-way stainless steel valves and requires user-supplied sources of zero air and
span gas.
Operating instructions for Automatic Zero/Span Check and Remote Zero/Span Check
are described in Sections 3.4 through 3.7.
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 76