Teledyne 300 Carbon Monoxide Alarm User Manual


 
1.5 Installation and overview
The Model 300 is shipped with the following standard equipment:
1. Power cord.
2. Instruction manual.
CAUTION
To avoid personal injury, always use two persons to lift and
carry the Model 300.
Upon receiving the Model 300 please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please
advise shipper first, then Teledyne API.)
2. When installing the Model 300, allow a minimum of 4 inches (100mm) of
clearance at the back of the instrument and 1 inch (25mm) of clearance on
each side for proper ventilation. Also, be sure that the clearance below the
chassis is unobstructed by at least the height of the instrument feet.
3. Connect sample inlet line to the sample port on rear panel.
NOTE
SEE FIGURE 1.4 FOR REAR PANEL PNEUMATIC
CONNECTIONS. SAMPLE GAS SHOULD ONLY COME
INTO CONTACT WITH PTFE, GLASS OR STAINLESS
STEEL. LEAK CHECK ALL FITTINGS WITH SOAP
SOLUTION.
MAXIMUM PRESSURE FOR LEAK CHECK IS 5 PSI.
CAUTION
Connect the exhaust fitting on the rear panel (See Fig. 1.4) to a
suitable vent outside the analyzer area.
4. Connect IZS inlet input (if installed) to a clean, dry air supply.
5. Connect a recording device to the terminal strip connections on the rear
panel (See Figure 1.2).
6. Connect the power cord to an appropriate power outlet (see the serial
number tag for correct voltage and frequency).
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 5