Teledyne 300 Carbon Monoxide Alarm User Manual


 
WARNING MESSAGES
Warning Message Meaning
SYSTEM RESET Issued whenever Analyzer is powered on
RAM INITIALIZED RAM was erased (incl. DAS reports)
SOURCE WARNING IR source < 2500 OR >= 5000 mV
BENCH HEAT SHUTDOWN Optical bench temp control not working
SAMPLE FLOW WARNING Sample flow < 500 cc/m or > 1000 cc/m
SAMPLE PRESSURE WARN Sample pressure < 15 or > 35 In-Hg-A
SAMPLE TEMP WARNING
Sample temperature < 10°C or > 50°C
BOX TEMP WARNING
Box temp. < 12°C or > 48°C
BENCH TEMP WARN
Optical Bench < 43°C or > 53°C
WHEEL TEMP WARN
Filter Wheel < 63°C or > 73°C
CANNOT DYN ZERO CO offset < -1500 or > +1500 mV
CANNOT DYN SPAN CO slope < 0.5 or > 2.0
V/F NOT INSTALLED A/D - I/O card not installed or bad
SYNC ERROR No modulation on detector output
TABLE 2.4
The menu field changes depending on the mode of the Analyzer and the buttons that have
been pressed. It indicates the current function of each of the 8 push buttons below the
display. See Section 2.2.2 for information on using the push buttons.
2.2.2 Programmable push buttons
The 8 push buttons below the display are programmable by the CPU in that their
functions change depending on the mode of the Analyzer or the operations being
performed. The legend above a button identifies its current function. If there is no
legend above a button, it has no function and will be ignored if pressed.
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 27