Teledyne 300 Carbon Monoxide Alarm User Manual


 
4. STARTING TIME: 23:30
5. DELTA DAYS: 7
6. DELTA TIME: 00:00
7. DURATION: 15
8. CALIBRATE: OFF
Example 4: to perform zero check once per day at 10:30 pm and a span check once per
week starting at 11:30 pm, 12/20/93.
1. Select any one of SEQx and program as example 1.
2. Select any other SEQx and program as example 3. Avoid setting
two or more sequences at the same time of the day.
3.5 Summary of front panel check and calibration
controls
The calibration controls are summarized below in terms of the button sequences used to
access them.
CALIBRATION CONTROLS
Button Sequence Function Default Limits
CALZ Begin zero check - -
CALZ-ZERO-ENTR Adjust CO conc. to zero - -
CALZ-EXIT Exit zero check - -
CALS Begin span check - -
CALS-CONC-ENTR Expected CO span value 40 PPM 1-1000 PPM
CALS-SPAN-ENTR Adj. CO conc. to span val. - -
CALS-EXIT Exit span check - -
CAL Begin M-P cal. - -
CAL-ZERO-ENTR Adj. CO conc. zero value - -
CAL-CONC-ENTR Expected CO span value 40 PPM 1-1000 PPM
CAL-SPAN-ENTR Adj. CO conc. to span val. - -
CAL-EXIT Exit M-P cal. - -
SETUP-ACAL-SEQ1 Setup auto-cal SEQ1 Disabled
SETUP-ACAL-SEQ2 Setup auto-cal SEQ2 Disabled
SETUP-ACAL-SEQ3 Setup auto-cal SEQ3 Disabled
TABLE 3.1
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 36