SETUP-RNGE-UNIT Set Measurement Units ppm ppb. ppm.
mg/m3,
ug/m3
SETUP-PASS Password Enable ON OFF-ON
SETUP-CLK-TIME Set Time-of-Day 00:00 00:00-23:59
SETUP-CLK-DATE Set Current Date 01 JAN 00 31 DEC 99
SETUP-MORE-COMM-BAUD RS-232 baud rate 2400 baud 300, 1200,
2400, 4800,
9600, 19.2
SETUP-MORE-COMM-ID Analyzer ID number 0000 0000-9999
SETUP-MORE-VARS-
DAS_HOLD_OFF
Set Hold-Off Interval 15 min 1 - 60 min
SETUP-MORE-VARS-
DYN_ZERO
Enable Remote
Dynamic Zero
Adjustment
OFF OFF-ON
SETUP-MORE-VARS-
DYN_SPAN
Enable Remote
Dynamic Span
Adjustment
OFF OFF-ON
SETUP-MORE-VARS-
RS232_MODE
Set RS-232 Mode 8
SETUP-MORE-VARS-
CLOCK_ADJ
Set Clock Adjustment
Rate
0 -15 to +15
SETUP-MORE-DIAG-DARK-
VIEW
View Dark Offset N/A N/A
SETUP-MORE-DIAG-
SIGNAL I/O
View the state of
internal signals
N/A N/A
SETUP-MORE-DIAG-
ANALOG OUTPUT
Generate Analog
Output Test Pattern
N/A N/A
SETUP-MORE-DIAG-
D/A CALIBRATION
Calibrate D/a\A and
A/D Converters
N/A N/A
SETUP-MORE-DIAG-
ELECTRICAL TEST
Generate Electrical
Test Output
N/A N/A
SETUP-MORE-DIAG-
DARK-CAL
Adjust Dark Offset 125mV 75-175mV
SETUP-MORE-DIAG-
TEST CHAN OUTPUT
Select TEST to Analog
Output
None
SETUP-MORE-DIAG-
RS-232 OUTPUT
Generate RS-232
Output Test Pattern
N/A N/A
TABLE 4.1
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 51